-- AndrejStuden - 03 Dec 2008

Serial readout, run 20081202

In serial mode, the resolution stays good:

testAm_U130_vpf711mV_thr14_serial.spectrumAligned.gif

The noise seen in sparse is due to common mode:

testAm_U130_vpf711mV_thr14_serial.commonMode.gif

There is a little change in gain (shift of mean for 0.01/400 ~ 0.5 ADC for peak at 250):

testAm_U130_vpf711mV_thr14_serialAndSparse.gainDistribution.gif

Equality of pedestals can be inferred from raw data comparison in serial/sparse mode. The square brackets show gaussian fit intervals & I cite the estimated peak. Error ~ 0.1 ADC count. This is for a typical channel. I didn't change the threshold!

testAm_U130_vpf711mV_thr14_serialAndSparse.spectrumRaw152.gif

Pedestals (in serial mode, N=1000 events)

testAm_U130_vpf711mV_thr14_serial.pedestal3.gif

Disribution of channels with respect to their pedestal (RMS = 9.8 ADC counts = 2.5 keV)

testAm_U130_vpf711mV_thr14_serial.pedestalDistribution3.gif